21 Code of Federal Regulations Parts 211 Subpart E-Control of Components and Drug Product Containers and Closures
21
Code of Federal Regulations
Parts
211
Subpart
E-Control of Components and Drug Product
Containers
and Closures
§
211.80 General requirements.
(a)
There shall be written procedures describing in sufficient detail the receipt,
identification, storage, handling, sampling, testing, and approval or rejection
of components and drug product containers and closures; such written procedures
shall be followed.
(b)
Components and drug product containers and closures shall at all times be
handled and stored in a manner to prevent contamination.
(c)
Bagged or boxed components of drug product containers, or closures shall be
stored off the floor and suitably spaced to permit cleaning and inspection.
(d)
Each container or grouping of containers for components or drug product
containers, or closures shall be identified with a distinctive code for each
lot in each shipment received. This code shall be used in recording the
disposition of each lot. Each lot shall be appropriately identified as to its
status (i.e., quarantined, approved, or rejected).
§ 211.82 Receipt and storage of untested components, drug product
containers, and closures.
(a) Upon receipt and before acceptance, each container or grouping
of containers of components, drug product containers, and closures shall be
examined visually for appropriate labeling as to contents, container damage or
broken seals, and contamination.
(b) Components, drug product containers, and closures shall be
stored under quarantine until they have been tested or examined, as
appropriate, and released. Storage within the area shall conform to the
requirements of §211.80.
§ 211.84 Testing and approval or rejection of components, drug
product containers, and
closures.
(a) Each lot of components, drug product containers, and closures
shall be withheld from use until the lot has been sampled, tested, or examined,
as appropriate, and released for use by the quality control unit.
(b) Representative samples of each shipment of each lot shall be
collected for testing or examination. The number of containers to be sampled,
and the amount of material to be taken from each container, shall be based upon
appropriate criteria such as statistical criteria for component variability,
confidence levels, and degree of precision desired, the past quality history of
the supplier, and the quantity needed for analysis and reserve where required
by § 211.170.
(c) Samples shall be collected in accordance with the following
procedures:
(1) The containers of components selected shall be cleaned where
necessary, by appropriate means.
(2) The containers shall be opened, sampled, and resealed in a
manner designed to prevent contamination of their contents and contamination of
other components, drug product containers, or closures.
(3) Sterile equipment and aseptic sampling techniques shall be
used when necessary.
(4) If it is necessary to sample a component from the top, middle,
and bottom of its container, such sample subdivisions shall not be composited
for testing.
(5) Sample containers shall be identified so that the following
information can be determined: name of the material sampled, the lot number,
the container from which the sample was taken, the date on which the sample was
taken, and the name of the person who collected the sample.
(6) Containers from which samples have been taken shall be marked
to show that samples have been removed from them.
(d) Samples shall be examined and tested as follows:
(1)
At least one test shall be conducted to verify the identity of each component
of a drug product. Specific identity tests, if they exist, shall be used.
(2)
Each component shall be tested for conformity with all appropriate written
specifications for purity, strength, and quality. In lieu of such testing by
the manufacturer, a report of analysis may be accepted from the supplier of a
component, provided that at least one specific identity test is conducted on
such component by the manufacturer, and provided that the manufacturer establishes
the reliability of the supplier's analyses through appropriate validation of
the supplier's test results at appropriate intervals.
(3)
Containers and closures shall be tested for conformance with all appropriate
written procedures. In lieu of such testing by the manufacturer, a certificate
of testing may be accepted from the supplier, provided that at least a visual
identification is conducted on such containers/closures by the manufacturer and
provided that the manufacturer establishes the reliability of the supplier's
test results through appropriate validation of the supplier's test results at
appropriate intervals.
(4)
When appropriate, components shall be microscopically examined.
(5)
Each lot of a component, drug product container, or closure that is liable to
contamination with filth, insect infestation, or other extraneous adulterant
shall be examined against established specifications for such contamination.
(6)
Each lot of a component, drug product container, or closure that is liable to
microbiological contamination that is objectionable in view of its intended use
shall be subjected to microbiological tests before use.
(e)
Any lot of components, drug product containers, or closures that meets the
appropriate written specifications of identity, strength, quality, and purity
and related tests under paragraph (d) of this section may be approved and
released for use. Any lot of such material that does not meet such specifications
shall be rejected.
§ 211.86 Use of approved components, drug product containers, and
closures.
Components, drug product containers, and closures approved for use
shall be rotated so that the oldest approved stock is used first. Deviation
from this requirement is permitted if such deviation is temporary and
appropriate.
§ 211.87 Retesting of approved components, drug product
containers, and closures.
Components, drug product containers, and closures shall be
retested or reexamined, as appropriate, for identity, strength, quality, and
purity and approved or rejected by the quality control unit in accordance with
§ 211.84 as necessary, e.g., after storage for long periods or after
exposure to air, heat or other conditions that might adversely affect the
component, drug product container, or closure.
§ 211.89 Rejected components, drug product containers, and
closures.
Rejected components, drug product containers, and closures shall
be identified and controlled under a quarantine system designed to prevent
their use in manufacturing or processing operations for which they are
unsuitable.
§ 211.94 Drug product containers and closures.
(a)
Drug product containers and closures shall not be reactive, additive, or
absorptive so as to alter the safety, identity, strength, quality, or purity of
the drug beyond the official or established requirements.
(b)
Container closure systems shall provide adequate protection against foreseeable
external factors in storage and use that can cause deterioration or
contamination of the drug product. (c) Drug product containers and closures
shall be clean and, where indicated by the nature of the drug, sterilized and
processed to remove pyrogenic properties to assure that they are suitable for
their intended use.
(d)
Standards or specifications, methods of testing, and, where indicated, methods
of cleaning, sterilizing, and processing to remove pyrogenic properties shall
be written and followed for drug product containers and closures.
Comments
Post a Comment